Title of article :
Disposable biosensor based on a hemoglobin colloidal gold-modified screen-printed electrode for determination of hydrogen peroxide
Author/Authors :
Ju، Huangxian نويسنده , , Xu، Xiaoxing نويسنده , , Liu، Songqin نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
A disposable reagentless hydrogen peroxide biosensor based on the direct electrochemistry of hemoglobin immobilized on a colloidal gold-modified screenprinted carbon electrode (Hb-Au-SPCE) was proposed. The electrochemical behavior of immobilized Hb at a SPCE was studied for the first time. The electrode reaction of immobilized Hb showed a surface-controlled process with an electron transfer rate constant of (0.40 + 0.02) s/sup -1/ determined in the scan rate range from 25 to 200 mV s/sup -1/. The Hb-Au-SPCE exhibited an electrocatalytic activity toward the reduction of hydrogen peroxide with a K/sub M//sup app/ value of 1.8 mM, which was allowed to be used as a disposable sensor for determination of hydrogen peroxide with a linear range from 1.0 * 10/sup -5/ M to 3.2 * 10/sup -4/ M, a detection limit of 5.5 * 10/sup -6/ M at 3*, a high sensitivity, fast response, and good selectivity, accuracy, and reproducibility. The disposable reagentless sensor was stable, low cost, and simple to use for detection of hydrogen peroxide in real samples.
Keywords :
Ethylene-Propylene Copolymer , liquid crystalline polymer , Microstructure , XRD , TGA , DSC , DMTA
Journal title :
IEEE Sensors Journal
Journal title :
IEEE Sensors Journal