Title of article :
Stability of ISFET and its new measurement protocol
Author/Authors :
Ito، نويسنده , , Yoshitaka، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
3
From page :
53
To page :
55
Abstract :
The aim of this report is to provide a new protocol for evaluating pH sensitivity films and stability of pH-ISFET. A light testing technique to evaluate stability of the pH sensing film was developed by using on–off lighting in a box. The long-term stability was strongly influenced by exposures to a fluorescent light and the drifts of ISFET were accelerated under the light with an intensity of more than 1000 lux. A good correlation for ISFET baseline drift was observed between conventional long term tests and this light testing technique.
Keywords :
ISFET , Stability of ISFET , Drift of ISFET