Title of article :
Spectroscopic hemoglobin and bilirubin measurement on optically opaque particulate fluids
Author/Authors :
Brandl، نويسنده , , Martin and Kellner، نويسنده , , Karlheinz and Posnicek، نويسنده , , Thomas and Bado، نويسنده , , Itka and Falkenhagen، نويسنده , , Dieter، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
7
From page :
711
To page :
717
Abstract :
Extracorporeal blood purification systems based on specific micro-adsorbents can be used for selective elimination of toxic substances from the patient. A suspension of blood plasma with micro-adsorbents is recirculated in the extracorporeal detoxification circuit. A safety system detecting erythrocyte damage by hemoglobin release as well as measuring the bilirubin elimination rate in the adsorbent circuit was developed. The sensor principle is based on broadband illumination and reflectivity measurement in the tubing where the adsorbent–plasma suspension is recirculated. From the specific reflection spectrum the hemoglobin and bilirubin content is calculated. Further, the effects of cross-sensitivity of hemoglobin and bilirubin on the detection accuracy are studied.
Keywords :
Blood purification , Hemoglobin sensor , Bilirubin sensor , optical sensors
Journal title :
Sensors and Actuators B: Chemical
Serial Year :
2013
Journal title :
Sensors and Actuators B: Chemical
Record number :
1442109
Link To Document :
بازگشت