Title of article :
Needle concentrator for gas chromatographic determination of BTEX in aqueous samples
Author/Authors :
Kubinec، نويسنده , , R?bert and Berezkin، نويسنده , , Victor G and G?rov?، نويسنده , , Ren?ta and Addov?، نويسنده , , Gabriela and Mra?nov?، نويسنده , , Helena and Soj?k، نويسنده , , Ladislav، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
A simple method of solventless extraction of volatile organic compounds (BTEX) from aqueous samples was developed and validated. A new arrangement of the full volume inside needle capillary adsorption trap (INCAT) device with Porapak Q as a sorbent material and wet alumina as a source of desorptive water vapour flow in a closed analytical system is presented. The analytical characteristics of developed device and of compared purge-and-trap (PTI) device for BTEX compounds are similar; the limits of detection as well as quantification are lower than 1 μg l−1.
Keywords :
Benzene , Ethylbenzene , Xylene , Toluene , Needle concentrator
Journal title :
Journal of Chromatography B
Journal title :
Journal of Chromatography B