Issue Information :
روزنامه با شماره پیاپی سال 2000
Pages :
7
From page :
151
To page :
157
Abstract :
An automatic method for the determination of metabolites of Ropivacaine in urine was set up. It utilizes supported liquid membrane extraction for sample clean-up and enrichment, followed by ion-pair chromatography determination using UV detection. The extraction was very selective with no observed interfering compounds from the urine matrix, permitting simple isocratic chromatographic analysis. The detection limits for spiked urine samples were 2–18 nM for the different compounds. The repeatability was 1–3% (RSD) with an internal standard that was also extracted, and about twice without this standard. A throughput of 3.3 samples per hour was achieved and the liquid membrane was stable for more than a week.
Journal title :
Journal of Chromatography B
Serial Year :
2000
Journal title :
Journal of Chromatography B
Record number :
1492618
Link To Document :
بازگشت