Abstract :
An in vitro microdialysis system was constructed for the measurement of catecholamines in pheochromocytoma cell culture medium. The novel microdialysis device is composed of a petri dish, a dialysis membrane and two transmission tubes. The dialysis membrane is located in the space of a petri dish such that it is immersed in the culture medium. Catecholamines contained in the culture medium diffused into a designed dialysis membrane with sufficient recovery (about 60%). Dialysates were collected by a sampling loop and introduced by an on-line injector to a microbore liquid chromatographic system for analysis of catecholamines. This assay yielded a detection limit of 0.2–0.5 pg/injection with acceptable intra- and inter-assay reproducibilities in 5 μl of dialysates. To evaluate the on-line microdialysis system, PC-12 cells were cultured in a petri dish within an incubator. The baseline concentration of dopamine in PC-12 cell culture medium was about 0.29 ng/ml which was elevated to 2.43 ng/ml after treatment with 0.5 mM potassium cyanide. In conclusion, the present microassay provides for the sensitive, direct measurement of catecholamines in culture medium while minimizing pretreatment procedures for sample preparation.