Title of article
Improvement of the high-speed log D assay using an injection marker for the water plug aspiration/injection method
Author/Authors
Nishimura، نويسنده , , Ikuko and Hirano، نويسنده , , Atsushi and Yamashita، نويسنده , , Taro and Fukami، نويسنده , , Takehiro، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
5
From page
2984
To page
2988
Abstract
An improved method for the high-speed assay for log D (HSLogD) employing an injection marker is described. The critical process of HSLogD is automated sampling of the water phase, which prevents contamination from the octanol phase. However, deployment of HSLogD sometimes produces incorrect and inaccurate values compared with data from closely related methods due to low solubility of compounds or instrument problems. Therefore, careful manual inspection of data is needed to confirm the results obtained by this method. To solve this problem, compounds were screened for suitability as injection markers and 2-(4-methoxyphenoxy)-4,8-dimethylquinoline was the most suitable compound examined. Introduction of the injection marker into the process successfully eliminated the need for the tedious and time-consuming manual inspection process.
Keywords
log , D measurements , High-speed log , d , Lipophilicity , Injection marker
Journal title
Journal of Chromatography A
Serial Year
2009
Journal title
Journal of Chromatography A
Record number
1511848
Link To Document