• Title of article

    Impurity profiling of acetylspiramycin by liquid chromatography–ion trap mass spectrometry

  • Author/Authors

    Wang، نويسنده , , Ming-juan and Pendela، نويسنده , , Murali and Hu، نويسنده , , Chang-qin and Jin، نويسنده , , Shao-hong and Hoogmartens، نويسنده , , Jos and Van Schepdael، نويسنده , , Ann and Adams، نويسنده , , Erwin، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    14
  • From page
    6531
  • To page
    6544
  • Abstract
    Investigation of acetylspiramycin (ASPM) and its related substances was carried out using a reversed-phase liquid chromatography/tandem mass spectrometry method. The identification of impurities in the ASPM complex was performed with a quadrupole ion trap mass spectrometer, with an electrospray ionization (ESI) source in the positive ion mode which provides MSn capability. A total of 83 compounds were characterized in commercial samples, among which 31 impurities that had never been reported and 31 partially characterized impurities were deduced using the collision-induced dissociation (CID) spectra of major ASPM components as templates. Most of the major impurities arise from the starting materials and the synthesis process. This work provides very useful information for quality control of ASPM and evaluation of its synthesis process.
  • Keywords
    Liquid chromatography/tandem mass spectrometry , 16-Membered macrolide , Impurity profile , Acetylspiramycin