• Title of article

    Voltammetry and determination of metronidazole at a carbon fiber microdisk electrode

  • Author/Authors

    Bartlett، نويسنده , , P.N. and Ghoneim، نويسنده , , E. and El-Hefnawy، نويسنده , , G. and El-Hallag، نويسنده , , I.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2005
  • Pages
    6
  • From page
    869
  • To page
    874
  • Abstract
    The electrochemistry of metronidazole, 1-(hydroxyethyl)-2-methyl-5-nitroimidazole, was investigated at a carbon fiber microdisk electrode in pH 9 Britton Robinson buffer. Under these conditions, the reduction of metronidazole is controlled by both mass transport to the microdisk and adsorption with an equilibrium constant of 4 × 103 mol−1 dm3 and a saturation coverage of 0.88 × 10−8 mol cm−2. The adsorption and accumulation of metronidazole on the surface of the carbon fiber allows its determination at low concentrations by square wave adsorptive stripping voltammetry. A detection limit for metronidazole of 5 × 10−7 mol dm−3 and a R.S.D. of 3.7% at 1 × 10−6 mol dm−3 (n = 4) were obtained with a two electrode system with no stirring during the accumulation step. Based on this method, a simple procedure for the determination of metronidazole in urine is described which requires no pre-treatment of the sample before analysis.
  • Keywords
    Carbon fiber microdisk , Metronidazole , Stripping voltammetry
  • Journal title
    Talanta
  • Serial Year
    2005
  • Journal title
    Talanta
  • Record number