Title of article :
A sensitive nonenzymatic hydrogen peroxide sensor using cadmium oxide nanoparticles/multiwall carbon nanotube modified glassy carbon electrode
Author/Authors :
Butwong، نويسنده , , Nutthaya and Zhou، نويسنده , , Lin and Ng-eontae، نويسنده , , Wittaya and Burakham، نويسنده , , Rodjana and Moore، نويسنده , , Eric and Srijaranai، نويسنده , , Supalax and Luong، نويسنده , , John H.T. and Glennon، نويسنده , , Jeremy D.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Abstract :
A simple sensing scheme for hydrogen peroxide has been described by electrochemical deposition of cadmium oxide (CdO) nanoparticles with ∼50 nm in diameter on a glassy carbon (GC) electrode modified with multiwall carbon nanotubes (MWCNT). The CdO/MWCNT modified sensor, with a surface coverage of 1.13 × 10−8 mol/cm2, displayed high synergistic electrocatalytic activity for H2O2. The sensor was capable of reducing H2O2 at −1.2 V vs. Ag/AgCl (H2O2 + 2H+ + 2e− → 2H2O) in a broad pH range. At the optimal pH 7, the sensor exhibited a detection limit of 0.1 μM, broad linearity from 0.5 to 200 μM, good reproducibility (RSD of 5.9%), and long-term stability. Uric acid, ascorbic acid and dopamine up to 100 μM provoked no signal response, attesting selectivity of the CdO/MWCNT modified electrode for hydrogen peroxide.
Keywords :
Multiwall carbon nanotube , Glassy carbon electrode , Hydrogen peroxide , Cadmium oxide
Journal title :
Journal of Electroanalytical Chemistry
Journal title :
Journal of Electroanalytical Chemistry