Title of article :
Rapid high-performance liquid chromatographic measurement of venlafaxine and O-desmethylvenlafaxine in human plasma: Application to management of acute intoxications
Author/Authors :
Matoga، نويسنده , , T and Pehourcq، نويسنده , , F and Titier، نويسنده , , Dumora، نويسنده , , F and Jarry، نويسنده , , C، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
6
From page :
213
To page :
218
Abstract :
Venlafaxine, a second-generation antidepressant, acts by inhibition of the reuptake of presynaptic noradrenaline and serotonin. The main metabolite, O-desmethylvenlafaxine was found biologically active. For toxicological purpose, a rapid specific and accurate RP-HPLC assay was developed for the simultaneous determination of venlafaxine and O-desmethylvenlafaxine in human plasma. A linear response was observed over the concentration range 0.2–4 μg/ml. A good accuracy (<8%) was achieved for all quality controls, with intra-day and inter-day variation coefficient less than 10%. Finally, no interference was observed with other psychotic drugs encountered in acute poisoning. This rapid method (run time <10 min) was used to manage four voluntary intoxications involving venlafaxine.
Keywords :
Venlafaxine , O-desmethylvenlafaxine
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Serial Year :
2001
Journal title :
Journal of Chromatography B Biomedical Sciences and Applications
Record number :
1705644