Title of article :
Mechanical evaluation of a respiratory device
Author/Authors :
Lima، نويسنده , , L.C. de and Duarte، نويسنده , , J.B.F. and Neto، نويسنده , , F.P. Lépore and Abe، نويسنده , , P.T. and Gastaldi، نويسنده , , A.C.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
This article aims to characterize the mechanical behaviour of the Flutter VRP1, a respiratory physiotherapy device designed to aid sputum clearance of the airways of patients. The device resembles a smoking pipe with a conical cavity containing a stainless steel sphere which floats up and down while the patient comes with a forced expiration through it. The sphereʹs oscillatory movement is function of the air flow rate and angular orientation of the device. When the sphereʹs oscillatory frequency matches the natural frequency of the patientʹs chestwall + abdomen system, it will produce resonance which, in turn, will enhance sputum clearance. A dynamical model of the Flutter was formulated and an experimental setup was assembled in order to study the oscillatory frequency of the sphere under different conditions of air flow rate, fluid pressure, angular orientation and sphereʹs material and weight. Interesting results presented by this article point to eventual mechanical optimization of the device and show information that could be beneficial to the professional of the respiratory physiotherapy.
Keywords :
Respiratory physiotherapy , Flutter VRP1 , Oscillatory frequency , Floating sphere
Journal title :
Medical Engineering and Physics
Journal title :
Medical Engineering and Physics