• Title of article

    Implantation-Related Complications of Implantable Cardioverter-Defibrillators and Cardiac Resynchronization Therapy Devices: A Systematic Review of Randomized Clinical Trials

  • Author/Authors

    van Rees، نويسنده , , Johannes B. and de Bie، نويسنده , , Mihلly K. and Thijssen، نويسنده , , Joep and Borleffs، نويسنده , , C. Jan Willem and Schalij، نويسنده , , Martin J. and van Erven، نويسنده , , Lieselot، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    6
  • From page
    995
  • To page
    1000
  • Abstract
    The number of implantable cardioverter-defibrillator (ICD) and cardiac resynchronization therapy (CRT) implantations is increasing drastically worldwide, and hence, the number of implanting centers is also increasing. Despite abundant data on the beneficial effect of these devices, little is known regarding safety and complication rates. Eleven ICD and 7 CRT trials were systematically reviewed to provide data on the frequency of in-hospital mortality and complications related to the implantation. Average in-hospital mortality was 2.7% in trials using both thoracotomy and nonthoracotomy ICDs, 0.2% in trials using nonthoracotomy ICDs, and 0.3% in CRT trials. The pneumothorax rate was similar between the nonthoracotomy ICD and CRT trials (0.9%) Coronary sinus complications occurred in 2.0% of patients undergoing CRT. Lead dislodgement rates were higher in CRT trials (5.7%) than in nonthoracotomy ICD trials (1.8%).
  • Keywords
    adverse events , Complications , cardiac resynchronization therapy , implantable cardioverter-defibrillator
  • Journal title
    JACC (Journal of the American College of Cardiology)
  • Serial Year
    2011
  • Journal title
    JACC (Journal of the American College of Cardiology)
  • Record number

    1752775