Title of article
Osmotic cavitation of elastomeric intraocular lenses
Author/Authors
Saylor، نويسنده , , David M. and Coleman Richardson، نويسنده , , D. and Dair، نويسنده , , Benita J. and Pollack، نويسنده , , Steven K.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2010
Pages
9
From page
1090
To page
1098
Abstract
In recent years, traditional rigid materials have been replaced with softer elastomers in intraocular lenses to minimize the size of the required surgical incision, thereby reducing patient recuperation time. However, water-filled cavities that may impact visual acuity are found in many of these new implants. We demonstrate that the cavitation observed in vivo can occur due to an osmotic pressure difference between the aqueous solution within the cavity and the external media in which the lens is immersed. By reducing the osmolarity of the external solution from 300 to 0 mM, we observe an increase in cavity volume of almost a factor of 30. Further, we have developed a model for cavity growth assuming the controlling factor is diffusion of hydrophilic molecules from the polymer matrix into the cavity. We find that the experimental observations are consistent with the model and suggest that oligomeric species generated during polymerization are responsible for the observed cavitation.
Keywords
intraocular lens (IOL) , elastomer , Osmotic pressure , Cavitation , diffusion
Journal title
Acta Biomaterialia
Serial Year
2010
Journal title
Acta Biomaterialia
Record number
1753669
Link To Document