Title of article :
Micro-solid phase extraction with liquid chromatography–tandem mass spectrometry for the determination of aflatoxins in coffee and malt beverage
Author/Authors :
Khayoon، نويسنده , , Wejdan Shakir and Saad، نويسنده , , Bahruddin and Salleh، نويسنده , , Baharuddin and Manaf، نويسنده , , Normaliza Hj Abdul and Latiff، نويسنده , , Aishah A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Abstract :
A single step extraction-cleanup procedure using porous membrane-protected micro-solid phase extraction (μ-SPE) in conjunction with liquid chromatography–tandem mass spectrometry for the extraction and determination of aflatoxins (AFs) B1, B2, G1 and G2 from food was successfully developed. After the extraction, AFs were desorbed from the μ-SPE device by ultrasonication using acetonitrile. The optimum extraction conditions were: sorbent material, C8; sorbent mass, 20 mg; extraction time, 90 min; stirring speed, 1000 rpm; sample volume, 10 mL; desorption solvent, acetonitrile; solvent volume, 350 μL and ultrasonication period, 25 min without salt addition. Under the optimum conditions, enrichment factor of 11, 9, 9 and 10 for AFG2, AFG1, AFB2 and AFB1, respectively were achieved. Good linearity and correlation coefficient was obtained over the concentration range of 0.4–50 ng g−1 (r2 0.9988–0.9999). Good recoveries for AFs ranging from 86.0–109% were obtained. The method was applied to 40 samples involving malt beverage (19) and canned coffee (21). No AFs were detected in the selected samples.
Keywords :
Aflatoxins , Micro-solid phase extraction , Liquid chromatography–tandem mass spectrometry , Mycotoxins , food
Journal title :
Food Chemistry
Journal title :
Food Chemistry