Title of article :
Selective identification and quantification of saccharin by liquid chromatography and fluorescence detection
Author/Authors :
Bruno، نويسنده , , Sergio N.F. and Cardoso، نويسنده , , Carlos R. and Maciel، نويسنده , , Mلrcia Mosca A. and Vokac، نويسنده , , Lidmila and da Silva Junior، نويسنده , , Ademلrio I. and Azevedo، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Pages :
7
From page :
309
To page :
315
Abstract :
High-pressure liquid chromatography with ultra-violet detection (HPLC–UV) is one of the most commonly used methods to identify and quantify saccharin in non-alcoholic beverages. However, due to the wide variety of interfering UV spectra in saccharin-containing beverage matrices, the same method cannot be used to measure this analyte accurately. We have developed a new, highly effective method to identify and quantify saccharin using HPLC with fluorescence detection (HPLC–FLD). The excitation wavelength (250 nm) and emission wavelength (440 nm) chosen increased selectivity for all matrices and ensured few changes were required in the mobile phase or other parameters. esence of saccharin in non-diet beverages – a fraud commonly used to replace more expensive sucrose – was confirmed by comparing coincident peaks as well as the emission spectra of standards and samples.
Keywords :
Fluorescence detection in liquid chromatography , Saccharin fraud , Selective saccharin determination
Journal title :
Food Chemistry
Serial Year :
2014
Journal title :
Food Chemistry
Record number :
1978185
Link To Document :
بازگشت