Author/Authors :
Sabini، نويسنده , , M.G and Bucciolini، نويسنده , , M and Cuttone، نويسنده , , G and Guasti، نويسنده , , A and Mazzocchi، نويسنده , , S and Raffaele، نويسنده , , L، نويسنده ,
Abstract :
In this work, the dose response of TLD-100 dosimeters has been studied in a 62 MeV clinical proton beams. The signal versus dose curve has been compared with the one measured in a 60Co beam. Different experiments have been performed in order to observe the thermal stress and the radiation damage effects on the detector sensitivity. A LET dependence of the TL response has been observed. In order to get a physical interpretation of these effects, a computerised glow-curve deconvolution has been employed. The results of all the performed experiments and deconvolutions are extensively reported, and the TLD-100 possible fields of application in the clinical proton dosimetry are discussed.