Title of article :
Spatial resolution of a GEM readout TPC using the charge dispersion signal
Author/Authors :
Boudjemline، نويسنده , , K. and Dixit، نويسنده , , M.S. and Martin، نويسنده , , J.-P. and Sachs، نويسنده , , K.، نويسنده ,
Pages :
6
From page :
22
To page :
27
Abstract :
A large volume Time Projection Chamber (TPC) is being considered for the central charged particle tracker for the detector for the proposed International Linear Collider (ILC). To meet the ILC-TPC spatial resolution challenge of ∼ 100 μ m with a manageable number of readout pads and channels of electronics, Micro Pattern Gas Detectors (MPGD) are being developed which could use pads comparable in width to the proportional-wire/cathode-pad TPC. We have built a prototype GEM readout TPC with 2 mm × 6 mm pads using the new concept of charge dispersion in MPGDs with a resistive anode. The dependence of transverse resolution on the drift distance has been measured for small angle tracks in cosmic ray tests without a magnetic field for Ar:CO2 (90:10). The GEM-TPC resolution with charge dispersion readout is significantly better than previous measurements carried out with conventional direct charge readout techniques.
Keywords :
Position-sensitive detectors , Micro-pattern Gas detectors , Gas electron multiplier , Gaseous detectors
Journal title :
Astroparticle Physics
Record number :
2028202
Link To Document :
بازگشت