Title of article :
Near-field scanning optical microscope probe analysis
Author/Authors :
Klapetek، نويسنده , , Petr and Bur??k، نويسنده , , Ji?? and Valtr، نويسنده , , Miroslav and Martinek، نويسنده , , Jan، نويسنده ,
Issue Information :
دوماهنامه با شماره پیاپی سال 2008
Pages :
6
From page :
671
To page :
676
Abstract :
In this article results of a comparison of two NSOM probe characterization methods are presented. Scanning electron microscopy analysis combined with electromagnetic field modeling using the finite difference in time domain method are compared with measured far-field radiation diagrams of NSOM probes. It is shown that measurement of far-field radiation diagrams can be an efficient tool for daily checking of the NSOM probes quality. Moreover, it is shown that the inner probe geometry has large influence on the directional radiation of an NSOM probe and the far-field radiation diagram can be used as a simple method to distinguish between different probe geometries.
Keywords :
Artifacts , NSOM
Journal title :
Ultramicroscopy
Serial Year :
2008
Journal title :
Ultramicroscopy
Record number :
2157193
Link To Document :
بازگشت