Title of article :
Optimizing the design of a moderator-based neutron detector for a flat response curve in the 2–14 MeV energy range
Author/Authors :
Hu، نويسنده , , Q.Y. and Li، نويسنده , , B.J. and Zhang، نويسنده , , D. and Guo، نويسنده , , H.S. and Yang، نويسنده , , G.Z. and Si، نويسنده , , F.N. and Liu، نويسنده , , J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
4
From page :
213
To page :
216
Abstract :
The efficiency of a neutron detector with boron trifluoride proportional tubes embedded in a polyethylene moderator was simulated with a Monte Carlo program. A moderator structure where the detector had uniform sensitivity for neutrons from 2 to 14 MeV was determined by simulation. A counter was built based on the simulation results. The counterʹs efficiencies were calibrated with an Am–Be source and an accelerator that served as a D–D and D–T neutron source. Experimental neutron efficiencies of these sources are approximately uniform. The simulated model was validated by the consistent results between the calculated and experimental data.
Keywords :
Energy response , Moderator , Monte Carlo simulation
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2009
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2211461