Title of article :
Adjusted blank correction method for UV–vis spectroscopic analysis of PTIO-coated filters used in nitrogen oxide passive samplers
Author/Authors :
DeForest Hauser، نويسنده , , Cindy and Battle، نويسنده , , Paul and Mace، نويسنده , , Nina، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
4
From page :
1823
To page :
1826
Abstract :
Passive sampling devices are popular in applications which do not require the monitoring of hourly concentrations. Nitrogen oxides are often collected using filters coated with 2-phenyl-4,4,5,5-tetramethylimidazoline-1-oxyl-3-oxide (PTIO). The filter extract can then be analyzed using flow injection analysis ion chromatography fitted with a copper/cadmium reduction column or UV–vis spectroscopy. When the latter is used to measure low concentrations of nitrogen oxides, absorbance by PTIO at the analytical wavelength of 545 nm contributes significantly. PTIO concentration on the filter also shows variation with filter storage and exposure time not accounted for in a single point blank subtraction at the analytical wavelength. A method is presented that uses a scaling factor to account for variations in concentration of PTIO on the field blank and provides a more accurate method for determining and correcting for the PTIO contribution to absorption when measuring ambient nitrogen oxide concentrations.
Keywords :
PTIO , UV–VIS , Passive samplers , NOX
Journal title :
Atmospheric Environment
Serial Year :
2009
Journal title :
Atmospheric Environment
Record number :
2234767
Link To Document :
بازگشت