• Title of article

    Diagnosing the cause of dynamic compression plate premature failure

  • Author/Authors

    Deen، نويسنده , , K.M. and Farooq، نويسنده , , A. and Sadiqui، نويسنده , , N.A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    8
  • From page
    350
  • To page
    357
  • Abstract
    The cause of premature failure of retrieved Dynamic Compression Plate (DCP) assembly from a 45 years old male was investigated by visual examination, chemical analysis of surface deposits by atomic absorption spectroscopy and Energy Dispersive spectroscopy coupled in a SEM. The electrochemical behavior of DCP and Screws was also evaluated in simulated body fluid through corrosion potential, Potentiodynamic Tafel scan and cyclic polarization. The chemical composition of DCP was not in conformity with ASTM F138-03. The higher concentration of ‘Ni’ and ‘Cr’ ions release and presence of chloride ions within the corrosion zone confirmed the dissolution of passive film by the aggressiveness of body environment. The negative corrosion potential, 2 times higher corrosion rate and varying passive current density of DCP than screws in simulated body fluid confirmed the higher susceptibility towards pitting and crevice corrosion. The higher protection tendency and breakdown potential with low passive current density of screws were also in support to lower corrosion tendency than DCP.
  • Keywords
    Crevice corrosion , pitting , electrochemical , Passive film , Stainless steel
  • Journal title
    Engineering Failure Analysis
  • Serial Year
    2013
  • Journal title
    Engineering Failure Analysis
  • Record number

    2339849