Title of article :
Measurement of beam power and profile for DNB on HT-7 tokamak
Author/Authors :
Liu، نويسنده , , Z.M. and Hu، نويسنده , , C.D. and Han، نويسنده , , X.P. and Li، نويسنده , , J. and Liu، نويسنده , , S. and Wan، نويسنده , , B.N.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
In normal experimental operation, a diagnostic neutral beam (DNB) can produce 6 A of extracted beam current in hydrogen at an energy of 49 keV with a pulse length of 100 ms. Hydrogen and deuterium beams can be produced as well. The diagnostic neutral beam has been added to the diagnostic set so that charge-exchange recombination spectroscopy (CXRS) can be used to acquire ion temperature and rotation. The beam power and beam profile distribution of the DNB injection can be obtained with a thermocouple probe measurement system on the HT-7 superconducting tokamak. The thermocouple probe measurement system with 13 thermocouples crossly distributed on the probe plate was used to measure the temperature rise of each coppery target, so the profile distribution of the ion/neutral beam was obtained by calculation. In this paper, the structure of the probe plate on the DNB for HT-7 tokamak and some measurement results are presented.
Keywords :
Diagnostic neutral beam , Target calorimeter system , Beams power , Beams profile distribution