• Title of article

    Determination of Bromethalin in Commercial Rodenticides Found in Consumer Product Samples by HPLC-UV-vis Spectrophotometry and HPLC-Negative-Ion APCI-MS

  • Author/Authors

    Mesmer، Mantai Z. نويسنده , , Flurer، Rick A. نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2001
  • Pages
    -48
  • From page
    49
  • To page
    0
  • Abstract
    A small amount of green particulate material is encountered in a consumer complaint sample. The green particulates in the sample are identified as a bromethalin-containing rodenticide using high-performance liquid chromatographic (HPLC)-UV-vis spectrophotometric and HPLC-negative-ion atmospheric pressure chemical ionization (APCI)-mass spectrometric (MS) approaches, which are commonly used for the detection and confirmation of bromethalin in grain-based rodenticides. The selective and sensitive nature of the MS detector makes it possible to determine bromethalin without extensive sample cleanup and preconcentration. The estimated detection limit with the UV-vis detector is 500 pg of bromethalin injected into the column. The extensive fragmentation of the bromethalin molecule under APCI conditions provides sufficient structural information for positive identification.
  • Keywords
    Fullerenes , Chemical synthesis , Infrared spectroscopy , Electronic paramagnetic resonance (EPR) , Organic compounds
  • Journal title
    JOURNAL OF CHROMATOGRAPHIC SCIENCE
  • Serial Year
    2001
  • Journal title
    JOURNAL OF CHROMATOGRAPHIC SCIENCE
  • Record number

    37744