Double-Helix Coil for Occlusion of Large Patent Ductus Arteriosus: Evaluation in Chronic Lamb Model
Author/Authors :
Ralph G. Grabitz MD، نويسنده , , Franz Freudenthal MD، نويسنده , , Matthias Sigler MD، نويسنده , , Trong-Phi Le MD، نويسنده , , Christoph Boosfeld PhD، نويسنده , , Stefan Handt MD، نويسنده , , G?tz von Bernuth MD، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
7
From page :
677
To page :
683
Abstract :
Objectives. We sought to evaluate the efficacy and tissue reaction of new miniature interventional device for occlusion of large patent ductus arteriosus (PDA) in neonatal lamb model. Background. variety of devices are used to close PDAs by interventional measures. Spring coils found to have high cumulative occlusion rate have thus far been limited to smaller PDAs because of the physical limitation of grip forces. Methods. Memory-shaped double-cone stainless steel coils with enhanced stiffness of the outer rings by double-helix configuration were mounted on titanium/nickel core wire. snap-in mechanism attaches the coil to the delivery wire, allowing intravascular coil retrieval and repositioning. The system was placed through 4F or 5F Teflon catheter. chronic lamb model (n = 8) of PD (>5 mm) was used in which ductus patency was secured by protocol of repetitive angioplasty procedures. The animals were killed after 1 to 181 days, and the ductal region was examined by inspection as well as by light and electron microscopy. Results. Placement of the coils within the PD was possible in all lambs. Before final detachment, the coils were retrieved or repositioned, or both, up to 12 times. In all but one animal the ductus was closed within 6 days after the procedure. The coils caused no infections or aortic and pulmonary artery obstruction. Histologic and electron microscopic studies revealed endothelial coverage of the implants but no foreign body reaction or local or systemic inflammation or erosion of the implant. Conclusions. The device effectively closed large PDAs in our model and may overcome the previous limitations of coils. Clinical trials are indicated.
Keywords :
Ansi , PDA , patent ductus arteriosus , American National Standards Institute
Journal title :
JACC (Journal of the American College of Cardiology)
Serial Year :
1998
Journal title :
JACC (Journal of the American College of Cardiology)
Record number :
480598
Link To Document :
بازگشت