Title of article :
Mandibular lengthening with an implanted motor-driven device: preliminary study in sheep
Author/Authors :
O. Ploder، نويسنده , , Ernst W. Mayr، نويسنده , , G. Schnetz، نويسنده , , E. Unger، نويسنده , , R. Ewers، نويسنده , , H. Plenk Jr، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Abstract :
Distraction osteogenesis with external or intraoral devices is an established method for lengthening human mandibles. In this preliminary study in three sheep a newly designed, fully implantable electromechanical device for mandibular lengthening was used. After osteotomy, the device was screw fixed to the mandible, and the power and control unit was inserted subcutaneously in the neck region. After a healing period of five days, the device was activated magnetically, allowing calibrated distraction steps of 0.04 mm/h, resulting in a total of 1.0 mm/day. Over a period of 14 days, a maximum mandibular lengthening of 13.6 mm could be achieved without transmucosal activation. Depending on stability of the screw fixation, membranous and/or cartilaginous bone formation was observed in the callus by radiological and histological evaluation. Further experimental research is ongoing to prove the clinical usefulness of this device.
Journal title :
British Journal of Oral and Maxillofacial surgery
Journal title :
British Journal of Oral and Maxillofacial surgery