Title of article :
Responses of alanine dosimeters to irradiations at cryogenic temperatures
Author/Authors :
F. Coninckx، نويسنده , , A. Janett، نويسنده , , T. Kojima، نويسنده , , S. Onori، نويسنده , , M. Pantaloni، نويسنده , , H. Sch?nbacher، نويسنده , , M. Tavlet، نويسنده , , A. Wieser، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
7
From page :
1223
To page :
1229
Abstract :
Present and future high-energy accelerators and particle detectors use and will use super-conducting technologies where components are submitted to radiation in a cryogenic environment. Radiation tests and dosimetry calibration have to be made under these conditions. Alanine-ESR dosimetry is world-wide recognised as a standard method for high-dose measurements at room-temperature irradiation. From earlier work, the system also seems suitable for low-temperature applications, despite a decrease of sensitivity. The aim of the present work was to estimate the temperature dependence of five different alanine based dosimeters. The dose-response functions at 290, 77 and 5 K were determined between 10 Gy and 80 kGy. All the results were consistent among the five laboratories: at a given dose, the amplitude of the ESR signal decreases with the irradiation temperature, moreover, this effect is found to be dose-dependent.
Journal title :
Applied Radiation and Isotopes
Serial Year :
1996
Journal title :
Applied Radiation and Isotopes
Record number :
539627