The standardization of 62Cu and experimental determinations of dose calibrator settings for generator-produced 62CuPTSM
Author/Authors
B. E. Zimmerman، نويسنده , , J. T. Cessna، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
12
From page
515
To page
526
Abstract
The Positron Emission Tomography agent 62Cu Pyruvaldehyde Bis(N4-methyl)thiosemicarbazone (PTSM) has been standardized by the National Institute of Standards and Technology (NIST) using 4πβ liquid scintillation counting with 3H-standard efficiency tracing. Using a measurement model developed at NIST for the determination of experimental dose calibrator dial settings for short-lived radionuclides, the correct dial settings for 62CuPTSM in a clinically useful geometry (a 35 ml plastic syringe containing 33 ml of solution) as well as the standard NIST 5 ml glass ampoule have been elucidated. This measurement model is fully described, as is the treatment of associated uncertainties. The correct instrument settings for the NIST Capintec CRC-12 re-entrant ionization chamber (‘dose calibrator’) were found to be ‘499±6’ and ‘489±8’, respectively, for the syringe and ampoule geometries. The stated uncertainties are expanded (k=2) uncertainties. The expanded uncertainties in the measured activity arising from these new dial settings are ±0.8% and ±1.4%, respectively, for the syringe and ampoule. The measured activities using these settings are lower than those obtained from the manufacture’s recommended setting of ‘448’ by 9.7% for the syringe and 7.7% for the ampoule, and underscore the need for experimental verification of the dose calibrator settings for each radionuclide and each geometry used in the clinic.
Journal title
Applied Radiation and Isotopes
Serial Year
1999
Journal title
Applied Radiation and Isotopes
Record number
540481
Link To Document