Title of article
Calibration and consistency of results of an ionization-chamber secondary standard measuring system for activity
Author/Authors
Heinrich Schrader، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
10
From page
325
To page
334
Abstract
Calibration in terms of activity of the ionization-chamber secondary standard measuring systems at the PTB is described. The measurement results of a Centronic IG12/A20, a Vinten ISOCAL IV and a radionuclide calibrator chamber for nuclear medicine applications are discussed, their energy-dependent efficiency curves established and the consistency checked using recently evaluated radionuclide decay data. Criteria for evaluating and transferring calibration factors (or efficiencies) are given.
Journal title
Applied Radiation and Isotopes