Title of article :
IAEA high-dose intercomparison in 60Co field
Author/Authors :
Kishor Mehta، نويسنده , , R. Girzikowsky، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
Nine high-dose calibration laboratories collaborated with the IAEA in a ‘double-blind’ intercomparison of absorbed dose in the 10–50 kGy dose range. Bureau International des Poids et Mesure undertook the analysis of the results as well as assisted in the development of the protocol. IAEA alanine-ESR transfer dosimeters were used for this intercomparison. The agreement was within 2.0% at 15 kGy and 2.2% at 45 kGy which is very good considering the variety of irradiation conditions employed by the participants. The mean of the dose values measured by the IAEA Dosimetry Laboratory was within 1% of the mean of the dose values stated by the participants for both dose values.
Journal title :
Applied Radiation and Isotopes
Journal title :
Applied Radiation and Isotopes