• Title of article

    Low level gamma spectrometry by beta–gamma coincidence

  • Author/Authors

    A. Luca، نويسنده , , P. de Felice، نويسنده , , G. Tanase، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    4
  • From page
    221
  • To page
    224
  • Abstract
    A beta–gamma coincidence system consisting of a 2π plastic beta detector and a Ge(Li) detector, is described. Such a system allows for a drastic reduction of background (not, vert, similar70 times) as well as of Minimum Detectable Activity (not, vert, similar7 times). An integral background (50–1500 keV) of less than 0.1 cps is obtained. The difficulty of the method, otherwise simple and inexpensive, is the precise determination of beta detector efficiency.
  • Keywords
    Gamma Spectrometry , Beta counters
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2000
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    540752