Abstract :
A method of preparation a seed-less active insert of iodine-125 ophthalmic applicator is described. Internal electrolysis was applied for fixing iodine-125 on the concave surface of the silver shell which is afterwards hermetically sealed inside a spherically shaped acrylic insert. The current–voltage characteristics of the galvanic cell used for deposition of silver iodide were determined. The results of the measurements of the energy spectra of the radiation emitted by a seed-less and seed-containing applicator are presented. The iodine-125 surface distribution uniformity on a silver shell was measured. The depth dose rate measurements indicate that the total activity incorporated in a seed-less applicator can be lower than that in seed-containing, while simultaneously assuring the desired dose rate.
Keywords :
Ophthalmic applicators , Iodine-125