    Independence of calibration curves for EBT Gafchromic films of the size of high-energy X-ray fields

  • Author/Authors

    Tsang Cheung، نويسنده , , Martin J. Butson، نويسنده , , Peter K.N. Yu، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    4
  • From page
    1027
  • To page
    1030
  • Abstract
    The EBT Gafchromic radiochromic film is a relatively new product designed specifically for dosimetry in radiation therapy. Due to the weak dependence of its response on the photon energy (variations are below 10% in the 50 kVp–10 MVp range), the film is ideal for dosimetry when the photon energy spectrum may be changing or unknown. In order to convert a map of optical densities into a map of absorbed radiation doses, a calibration curve constructed on the basis of standard calibration films is necessary. Our results have shown that, with the EBT Gafchromic film, one can use the same calibration curve for 6-MV X-ray fields of any size in the range from 5×5 cm2 up to 40×40 cm2. This is not the case for radiographic films, such as Kodak X-Omat V, whose response to the same dose varies approximately by 10% depending on the field size in this range. This insensitivity of the EBT Gafchromic film to size of the radiation field makes it possible to assess doses delivered by small radiation fields. With the help of this film, it was shown that the output factor for a 0.5×0.5 cm2 field is 0.60±0.03 (2SD) relative to the 10×10 cm2 field.
  • Keywords
    radiotherapy , Gafchromic , radiochromic film , EBT , energy dependence , Field size , scatter , calibration
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2006
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    548014