• Title of article

    Investigating the background of a 1-cm3 quartz gas proportional counter

  • Author/Authors

    George A. Klouda، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    5
  • From page
    860
  • To page
    864
  • Abstract
    The background count rate of a 1-cm3 quartz gas proportional counter (GPC) was investigated for measuring natural radioactivity levels of small samples. Using a new multiparameter digital counting system, spurious or afterpulses that contribute to background were identified by their pulse waveform and by a special event indicator defined by the time (<2000 μs) between two successive pulses. With the GPC in coincidence with a low-background NaI(Tl) scintillation guard detector, the weighted mean and standard uncertainty anticoincidence counts per day (d−1) was 6.2±0.4 d−1.
  • Keywords
    Gas proportional counter , background , Risetime , Afterpulse , Digital acquisition , Low-level decay counting , Pulse waveform , Natural radioactivity
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2008
  • Journal title
    Applied Radiation and Isotopes
  • Record number