• Title of article

    Surface and build-up region dose analysis for clinical radiotherapy photon beams

  • Author/Authors

    E. Ishmael Parsai، نويسنده , , Diana Shvydka، نويسنده , , David Pearson، نويسنده , , M. Gopalakrishnan، نويسنده , , John J. Feldmeier، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    5
  • From page
    1438
  • To page
    1442
  • Abstract
    The standard clinical approach of dose measurement using a Farmer type fixed plane parallel cylindrical ionization chamber produces erroneous results in the build-up region. We studied dose distribution in this region using a Monte Carlo simulation technique and compared our results with data measured using extrapolation, parallel plate, and cylindrical farmer type ionization chambers for 6 and 10 MV photon beams from two different accelerators. The extrapolation chamber data agreed favorably with the Monte Carlo results, suggesting that dose at the skin surface and a few mm beneath is significantly lower than conventionally accepted values.
  • Keywords
    Monte Carlo simulations , Extrapolation chamber , surface dose , Build-up region
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2008
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    548557