• Title of article

    Determination of the diffuser reference plane for accurate illuminance responsivity calibrations

  • Author/Authors

    Hovila، Jari نويسنده , , Mustonen، Maria نويسنده , , Karha، Petri نويسنده , , Ikonen، Erkki نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    -5893
  • From page
    5894
  • To page
    0
  • Abstract
    It is difficult to predict where the effective measurement plane is situated with dome-shaped diffusers often used in commercial photometers and radiometers. Insufficient knowledge of this plane could lead to large systematic errors in calibration of the illuminance responsivity of photometers. We propose a method that can be used to determine this reference plane accurately, based on the inverse-square law between the measured signal and the distance from the source. The method is demonstrated with three commercial photometers with dome-shaped diffusers of different geometries. By taking into account the measured shifts of the reference planes (5.0+-0.5 mm, 7.8+-0.3 mm, and 8.5+-0.7 mm), we reduced the systematic measurement errors up to 2% to statistical uncertainty components at the level of 0.2%.
  • Keywords
    instrumentation , Metrology , Optical devices , Diffusers , Measurement , photometry
  • Journal title
    Applied Optics
  • Serial Year
    2005
  • Journal title
    Applied Optics
  • Record number