چكيده لاتين :
Radiation therapy requires precision to avoid unintended irradiation of normal organs. Electronic Portal Imaging
Devices (EPIDs), can help with precise patient positioning for accurate treatment. EPIDs are now bundled with new
linear accelerators, or they can be purchased from the Linac manufacturer for retrofit. Retrofitting a third party EPID to a
linear accelerator can pose challenges. The authors describe a relatively inexpensive third party CCD camera-based
EPID manufactured by TheraView (Cablon Medical B.V.), installed onto a Siemens Primus linear accelerator, and
integrated with a Lantis record and verify system, an Oldelft simulator with Digital Therapy Imaging (DTI) unit, and a
Philips ADAC Pinnacle treatment planning system (TPS). This system integrates well with existing equipment and its
software can process DICOM images from other sources. The system provides a complete imaging system that
eliminates the need for separate software for portal image viewing, interpretation, analysis, archiving, image guided
radiation therapy and other image management applications. It can also be accessed remotely via safe VPN tunnels.
TheraView EPID retrofit therefore presents an example of a less expensive alternative to linear accelerator
manufacturers’ proprietary EPIDs suitable for implementation in third world countries radiation therapy departments
which are often faced with limited financial resources.