• DocumentCode
    1008732
  • Title

    The SISWICH, a detector telescope with intrinsic calibration

  • Author

    Friese, J. ; Gillitzer, A. ; Körner, H.J. ; Reinhold, J. ; Peter, M. ; Maier, M.R.

  • Author_Institution
    Tech. Univ. Munchen, Germany
  • Volume
    40
  • Issue
    4
  • fYear
    1993
  • fDate
    8/1/1993 12:00:00 AM
  • Firstpage
    443
  • Lastpage
    446
  • Abstract
    Phoswich counters provide adequate energy resolution and particle identification up to atomic numbers Z=20, but calibration has proved difficult. To overcome this difficulty, a silicon-scintillator sandwich (SISWICH) has been built. This detector consists of a Si PIN photodiode glued to the front of a CsI(Tl) scintillator crystal. The signal coming from the PIN diode contains the energy deposited by the particle by direct ionization in the silicon, which is very fast, and the light output of the scintillator, which is slow. The direct component can be calibrated against a pulser. Since the energy loss of a particular particle has been measured, its energy is known, and this can be used to calibrate the signal from the scintillator
  • Keywords
    calibration; ionisation chambers; scintillation counters; semiconductor counters; CsI(Tl); CsI:Tl; SISWICH; Si; Si PIN photodiode; calibration; energy loss; light output; Calibration; Energy loss; Energy measurement; Energy resolution; Ionization; Loss measurement; PIN photodiodes; Radiation detectors; Silicon; Telescopes;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.256597
  • Filename
    256597