• DocumentCode
    1010120
  • Title

    An elliptical orbit backprojection filtering algorithm for SPECT

  • Author

    Gullberg, Grant T. ; Zeng, Gengsheng L.

  • Author_Institution
    Dept. of Radiol., Utah Univ., Salt Lake City, UT, USA
  • Volume
    40
  • Issue
    4
  • fYear
    1993
  • fDate
    8/1/1993 12:00:00 AM
  • Firstpage
    1102
  • Lastpage
    1106
  • Abstract
    An orbit geometry where the central projection ray of the detector is not constrained to pass through the center-of-rotation for all projection angles can be useful to reduce uniformity artifacts and to avoid truncation when imaging noncentrally located organs. A backprojection filtering algorithm is presented for parallel geometry with an elliptical orbit acquisition. The algorithm is verified by computer simulations. It is shown that the point response function is shift invariant and has an analytical expression which is not the common 1/r but is a function of the orbit geometry
  • Keywords
    computerised tomography; data acquisition; medical image processing; patient diagnosis; radioisotope scanning and imaging; SPECT; center-of-rotation; central projection ray; elliptical orbit acquisition; elliptical orbit backprojection filtering algorithm; noncentrally located organs; orbit geometry; point response function; projection angles; truncation; uniformity artifacts; Cameras; Computer simulation; Face detection; Filtering algorithms; Gamma ray detection; Gamma ray detectors; Geometry; Image reconstruction; Optical collimators; Tomography;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.256719
  • Filename
    256719