Abstract :
An orbit geometry where the central projection ray of the detector is not constrained to pass through the center-of-rotation for all projection angles can be useful to reduce uniformity artifacts and to avoid truncation when imaging noncentrally located organs. A backprojection filtering algorithm is presented for parallel geometry with an elliptical orbit acquisition. The algorithm is verified by computer simulations. It is shown that the point response function is shift invariant and has an analytical expression which is not the common 1/r but is a function of the orbit geometry
Keywords :
computerised tomography; data acquisition; medical image processing; patient diagnosis; radioisotope scanning and imaging; SPECT; center-of-rotation; central projection ray; elliptical orbit acquisition; elliptical orbit backprojection filtering algorithm; noncentrally located organs; orbit geometry; point response function; projection angles; truncation; uniformity artifacts; Cameras; Computer simulation; Face detection; Filtering algorithms; Gamma ray detection; Gamma ray detectors; Geometry; Image reconstruction; Optical collimators; Tomography;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.256719
Filename :
256719
Link To Document :
بازگشت