• DocumentCode
    1010262
  • Title

    Accurate intensity correction for endorectal surface coil MR imaging of the prostate

  • Author

    Singh, Manbir ; NessAiver, Moriel

  • Author_Institution
    Dept. of Radiol. & Biomed. Eng., Univ. of Southern California, Los Angeles, CA, USA
  • Volume
    40
  • Issue
    4
  • fYear
    1993
  • fDate
    8/1/1993 12:00:00 AM
  • Firstpage
    1169
  • Lastpage
    1173
  • Abstract
    An accurate technique is developed to correct the intensity variation in the magnetic resonance images acquired with an endorectal surface coil for prostate imaging. The technique relies on reconstructing the in vivo size, shape, and location of the coil during imaging, wherefrom a pixel-by-pixel correction is applied to the coil´s sensitivity profile in accordance with the Biot-Savart law. The method is evaluated by a test-object containing multiple layers of beads embedded in gelatine. Qualitatively, the improvement in visualizing regions of interest after corrections is obvious. Quantitatively, the local SNR shows an improvement by a factor of 1.5-2. Several human studies are performed. The technique shows remarkable improvement in visualizing the prostate region
  • Keywords
    biomedical NMR; medical image processing; patient diagnosis; Biot-Savart law; endorectal surface coil; in vivo size; magnetic resonance images; pixel-by-pixel correction; prostate imaging; Coils; Image reconstruction; In vivo; Magnetic resonance; Magnetic resonance imaging; Pixel; Shape; Surface reconstruction; Testing; Visualization;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.256731