Performance studies of a multiwire chamber, with 1 mm anode wire spacing, to be used for 2-D position-sensitive detection of X-rays
Author
Mathieson, E. ; Smith, G.C.
Author_Institution
Leicester Univ., UK
Volume
37
Issue
2
fYear
1990
fDate
4/1/1990 12:00:00 AM
Firstpage
187
Lastpage
189
Abstract
Experimental and theoretical studies have been made of the performance of a multiwire chamber, with 1-mm anode wire spacing, to be used for position-sensitive detection of X-rays. The chamber has a wire upper cathode, with the wires parallel to the anode wires. The studies have shown that, from considerations of the spacing and registration of the cathode wires, some useful improvements in performance can be obtained. These improvements concern both position modulation and gain modulation, normal to the anode wire direction. The fine system presented has been shown to achieve, for 1-mm wire spacing, a resolution of 125 μm FWHM (full width at half maximum) at 5.4-keV X-ray energy
Keywords
X-ray detection and measurement; position sensitive particle detectors; proportional counters; 2-D position-sensitive detection; FWHM; X-ray energy; anode wire spacing; gain modulation; multiwire chamber; position modulation; wire upper cathode; Anodes; Cathodes; Geometry; Laboratories; Optical imaging; Position sensitive particle detectors; Synchrotrons; Wire; X-ray detection; X-ray detectors;