DocumentCode
1041786
Title
Benchmarking the Sandia Pulsed Reactor III cavity neutron spectrum for electronic parts calibration and testing
Author
Kelly, John G. ; Griffin, Patrick J. ; Fan, Wesley C.
Author_Institution
Sandia Nat. Lab., Albuquerque, NM, USA
Volume
40
Issue
6
fYear
1993
fDate
12/1/1993 12:00:00 AM
Firstpage
1418
Lastpage
1425
Abstract
The SPR (Sandia Pulsed Reactor) III bare cavity spectrum and integral parameters have been determined with 24 measured spectrum sensor responses and an independent, detailed, transport calculation. This environment qualifies as a benchmark field for electronic parts testing. The measured SPR III cavity spectrum is characterized by a very smooth curve that predicts the measured activities with a standard deviation of 3.5% for the 24 sensors. Within the limitations of the computational model, the spectrum is also very close to that provided by MCNP (Monte Carlo Neutron Photon) calculations
Keywords
calibration; electron device testing; electronic equipment testing; measurement standards; neutron effects; radiation hardening (electronics); test facilities; Monte Carlo calculation; Sandia Pulsed Reactor III; benchmark field; cavity neutron spectrum; electronic parts calibration; electronic parts testing; radiation hardness; spectrum sensor responses; transport calculation; Benchmark testing; Computational modeling; Electronic equipment testing; Inductors; Measurement standards; Monte Carlo methods; Neutrons; Optical computing; Pulse measurements; Sensor phenomena and characterization;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.273523
Filename
273523
Link To Document