• DocumentCode
    1065340
  • Title

    Simplified dynamic focus method for time domain OCT

  • Author

    Hughes, M. ; Podoleanu, A. Gh

  • Author_Institution
    Sch. of Phys. Sci., Univ. of Kent, Canterbury
  • Volume
    45
  • Issue
    12
  • fYear
    2009
  • Firstpage
    623
  • Lastpage
    624
  • Abstract
    A new optical arrangement for performing dynamic focus in time domain optical coherence tomography (OCT) is demonstrated. Unlike previously reported schemes which require mechanical coupling of the object and reference arms, this method is confined to the object arm only and therefore does not impose design constraints on the OCT system layout. The scheme is tested on a high lateral resolution OCT system (NA = 0.13) and it is shown that the effective depth of focus is extended from 200 mum to better than 2 mm. The optimum correction is for media with a mean refractive index of 1.4.
  • Keywords
    image resolution; optical design techniques; optical focusing; optical tomography; refractive index; dynamic focus method; high-lateral resolution OCT system layout design; optical coherence tomography; refractive index; time domain OCT;
  • fLanguage
    English
  • Journal_Title
    Electronics Letters
  • Publisher
    iet
  • ISSN
    0013-5194
  • Type

    jour

  • DOI
    10.1049/el.2009.0672
  • Filename
    5069775