Simplified dynamic focus method for time domain OCT
Author
Hughes, M. ; Podoleanu, A. Gh
Author_Institution
Sch. of Phys. Sci., Univ. of Kent, Canterbury
Volume
45
Issue
12
fYear
2009
Firstpage
623
Lastpage
624
Abstract
A new optical arrangement for performing dynamic focus in time domain optical coherence tomography (OCT) is demonstrated. Unlike previously reported schemes which require mechanical coupling of the object and reference arms, this method is confined to the object arm only and therefore does not impose design constraints on the OCT system layout. The scheme is tested on a high lateral resolution OCT system (NA = 0.13) and it is shown that the effective depth of focus is extended from 200 mum to better than 2 mm. The optimum correction is for media with a mean refractive index of 1.4.
Keywords
image resolution; optical design techniques; optical focusing; optical tomography; refractive index; dynamic focus method; high-lateral resolution OCT system layout design; optical coherence tomography; refractive index; time domain OCT;
fLanguage
English
Journal_Title
Electronics Letters
Publisher
iet
ISSN
0013-5194
Type
jour
DOI
10.1049/el.2009.0672