Abstract :
X-ray detector systems are powerful tools: in combination with tomographic methods they provide volumetric data of samples in a non-destructive manner which is of high interest for, e.g., biology, medicine or materials research. The detector able to provide images with submicrometer spatial resolution frequently consists of a scintillator screen, light microscopy optics and a digital camera. Here, the scintillator converts the X-rays into a visible light image which is projected onto the camera by the light optics. In order to perform high resolution imaging Single Crystal Film (SCF) scintillators 1 mum to 30 mum thin are required due to the limited depth of focus of the microscopy optics. Thin SCFs can be obtained via liquid phase epitaxy (LPE). A drawback is that a detector working with SCFs suffers from low efficiency (2% at 50 keV) owing to their limited thickness. The detective quantum efficiency (DQE) is here mainly limited by the low absorption of X-rays and the light yield in the thin scintillator layer. Performances, i.e absorption, light yield, afterglow of operational systems at the European Synchrotron Radiation Facility (ESRF) using YAG:Ce (Y3Al5O12:Ce), LAG:Eu (Lu3Al5O12:Eu) and GGG:Eu (Gd3Ga5O12:Eu) scintillators will be presented and compared to new LSO:Tb (Lu2SiO5:Tb) scintillators developed in the framework of an European project , . A new concept to improve the efficiency of detection in the 20 keV - 40 keV energy range with 1 mum spatial resolution will be presented. This concept based on multilayer scintillators is realised by the LPE process as well. First results will be illustrated with X-ray images and will demonstrate the absorption efficiency improvement of the X-ray detector. The expected performance is 7 times better than the LAG-based scintillators.
Keywords :
X-ray absorption; X-ray apparatus; X-ray detection; X-ray microscopy; cameras; position sensitive particle detectors; solid scintillation detectors; thin film devices; GGG:Eu scintillators; Gd3Ga5O12:Eu; LAG:Eu scintillators; LSO:Tb scintillators; Lu2SiO5:Tb; Lu3Al5O12:Eu; X-ray absorption; X-ray detector systems; Y3Al5O12:Ce; YAG:Ce scintillators; afterglow; detective quantum efficiency; digital camera; electron volt energy 20 keV to 40 keV; electron volt energy 50 keV; light microscopy optics; light yield; liquid phase epitaxy; scintillator screen; single crystal film scintillator; size 1 mum to 30 mum; synchrotron-based hard X-ray microimaging; tomographic methods; visible light image; Biomedical imaging; Biomedical optical imaging; Electromagnetic wave absorption; Optical films; Optical microscopy; Spatial resolution; Systems biology; Tomography; X-ray detectors; X-ray imaging; LSO:Tb; X-ray detector; luminescence; micro-imaging; micro-tomography; submicrometer resolution; synchrotron instrumentation; thin screen scintillator;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2009.2015878
Filename :
5076104
Link To Document :
بازگشت