• DocumentCode
    1095705
  • Title

    An International Comparison of Frequency by Means of a Luminous Quartz Resonator

  • Author

    Jimbo, S.

  • Author_Institution
    Electrotechnical Laboratory, Tokyo, Japan
  • Volume
    18
  • Issue
    11
  • fYear
    1930
  • Firstpage
    1930
  • Lastpage
    1934
  • Abstract
    The international comparison of frequency standards made with the luminous quartz resonator shows the different laboratories,-Physikalisch-technische Reichsanstalt, National Physical Laboratory, Bureau of Standards, and Electrotechnical Laboratory,-to be in agreement to one part in 105when used to calibrate the resonator at its flexuralfundamental of about 10 kc, due allowance being madefor the temperature coefficient of the resonator in this mode, namely, about 1 part in 105and negative. The observed agreement seems limited by the luminous glow resonator used rather than by any difference between the laboratory standards compared.
  • Keywords
    Coils; Electrodes; Frequency; Inductance; Laboratories; Optical resonators; Resonance; Temperature; Yarn;