  • Abstract
    The performance of a xenon-filled gas proportional-scintillation counter is reported, where the VUV Xe scintillation is detected by KI or KBr photocathodes on a microstrip plate. The detector energy resolution R and gain G for 5.9 keV X-rays are presented and compared with previous results obtained with Csl photocathodes. R and G are studied as a function of the halide film thickness and scintillation region length. The best energy resolutions reached values of 12% and 15% for KI and KBr photocathodes, respectively.
  • Keywords
    gas scintillation detectors; photocathodes; position sensitive particle detectors; proportional counters; Csl photocathodes; KBr photocathode; KBr-covered microstrip plate; KI photocathode; KI-covered microstrip plate; VUV Xe scintillation; X-rays; halide film thickness; photosensor; scintillation region length; xenon-filled gas proportional-scintillation counter; Backscatter; Cathodes; Energy resolution; Feedback; Microstrip; Radiation detectors; Scintillation counters; Solid scintillation detectors; X-ray detection; X-ray detectors; CsI; KBr; KI; X-ray detector; gas proportional scintillation counter (GPSC); microstrip plate; photosensor;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2007.902357
  • Filename
    4291711