• DocumentCode
    1104559
  • Title

    Individual characterization of broadband sampling oscilloscopes with a nose-to-nose calibration procedure

  • Author

    Verspecht, Jan ; Rush, Ken

  • Author_Institution
    ELEC, Vrije Univ., Brussels, Belgium
  • Volume
    43
  • Issue
    2
  • fYear
    1994
  • fDate
    4/1/1994 12:00:00 AM
  • Firstpage
    347
  • Lastpage
    354
  • Abstract
    A method is proposed to find the individual impulse response of broad-band sampling oscilloscopes (HP54124T, 50 GHz bandwidth). The method is based on the fact that, when the oscilloscope is sampling a dc voltage, pulses are launched from the sampler towards the input connector. These pulses contain information on the scope´s characteristic and can be measured by a second oscilloscope. This type of measurement is called a “nose-to-nose” calibration. Applying deconvolution techniques to the result of this measurement, the characteristic of the two scopes can be found, assuming that the two scopes are identical. To avoid relying on this assumption, three oscilloscopes can be used. The nose-to-nose is then applied three times, with a different pair of scopes connected together each time. The individual characteristics of the three sampling oscilloscopes are then calculated. Both SPICE simulations and close correspondence between swept-sine measurements and this method indicate that it is probably the most accurate method available at this moment to calibrate broadband sampling oscilloscopes. Several measurement uncertainties and practical problems are identified. The phase contribution of the sampling aperture can never be determined by this method, but there is strong evidence that the effect is negligible. Practical measurement problems are related to linearity of the samplers, additive noise, time-base jitter, and time-base drift
  • Keywords
    SPICE; calibration; characteristics measurement; computerised instrumentation; digital simulation; oscilloscopes; signal processing; 50 GHz; HP54124T; SPICE simulation; additive noise; broadband sampling oscilloscopes; close correspondence; dc voltage; deconvolution; impulse response; linearity; measurement uncertainties; modelling errors; nose-to-nose calibration; sampling aperture; swept-sine measurement; time-base drift; time-base jitter; Bandwidth; Calibration; Connectors; Deconvolution; Measurement uncertainty; Oscilloscopes; Pulse measurements; SPICE; Sampling methods; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.293446
  • Filename
    293446