• DocumentCode
    112121
  • Title

    1D-3D Registration for Intra-Operative Nuclear Imaging in Radio-Guided Surgery

  • Author

    Vetter, Christoph ; Lasser, Tobias ; Okur, Asli ; Navab, Nassir

  • Author_Institution
    Dept. for Comput. Aided Med. Procedures (CAMP), Tech. Univ. MunchenMünchen, Garching, Germany
  • Volume
    34
  • Issue
    2
  • fYear
    2015
  • fDate
    Feb. 2015
  • Firstpage
    608
  • Lastpage
    617
  • Abstract
    3D functional nuclear imaging modalities like SPECT or PET provide valuable information, as small structures can be marked with radioactive tracers to be localized before surgery. This positional information is valuable during surgery as well, for example when locating potentially cancerous lymph nodes in the case of breast cancer. However, the volumetric information provided by pre-operative SPECT scans loses validity quickly due to posture changes and manipulation of the soft tissue during surgery. During the intervention, the surgeon has to rely on the acoustic feedback provided by handheld gamma-detectors in order to localize the marked structures. In this paper, we present a method that allows updating the pre-operative image with a very limited number of tracked readings. A previously acquired 3D functional volume serves as prior knowledge and a limited number of new 1D detector readings is used in order to update the prior knowledge. This update is performed by a 1D-3D registration algorithm that registers the volume to the detector readings. This enables the rapid update of the visual guidance provided to the surgeon during a radio-guided surgery without slowing down the surgical workflow. We evaluate the performance of this approach using Monte-Carlo simulations, phantom experiments and patient data, resulting in a positional error of less than 8 mm which is acceptable for surgery. The 1D-3D registration is also compared to a volumetric reconstruction using the tracked detector measurements without taking prior information into account, and achieves a comparable accuracy with significantly less measurements.
  • Keywords
    Monte Carlo methods; image reconstruction; image registration; medical image processing; phantoms; single photon emission computed tomography; surgery; 1D-3D registration; 3D functional nuclear imaging modalities; Monte-Carlo simulations; PET; SPECT; intraoperative nuclear imaging; phantom experiments; radioguided surgery; volumetric reconstruction; Detectors; Lymph nodes; Phantoms; Single photon emission computed tomography; Surgery; Visualization; Volume measurement; Image reconstruction; nuclear imaging; registration; single photon emission computed tomography;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.2014.2363551
  • Filename
    6926859