Effect of Electrode Surface Area on Thresholds for AC Stimulation and Ventricular Fibrillation
Author :
Patel, Shivani Govind ; Malkin, Robert A.
Author_Institution :
John H. Stroger Hospital of Cook County, Chicago
Volume :
54
Issue :
10
fYear :
2007
Firstpage :
1829
Lastpage :
1836
Abstract :
Unintended, weak AC stimulation (leakage currents) from medical devices can cause blood pressure collapse and ventricular fibrillation (VF), potentially even death. Yet, little is understood about AC cardiac stimulation. The objective of this paper is to establish the relationship between the stimulation and VF thresholds for electrode size and stimulation frequency. Twenty- four retired male breeder guinea pigs were anesthetized with isoflurane, a tracheotomy and thoracotomy were performed, and vitals were monitored using the lead II ECG and an optical plethysmograph. The circular flat ends of eleven stainless steel rods were used as electrodes with areas ranging from 0.1 to 26.79 mm2. In the first study, 60-Hz AC stimuli of 5 s duration were delivered with strengths from 25-3000 muA or until VF was induced. In the second group, the current thresholds at 20, 40, 80, and 160 Hz were determined at electrode areas of 0.2,2.01, and 16.4 mm2. Reactions were categorized as having no effect, having some effect (EFFECT, typically blood pressure collapse), and inducing VF. On a log-log scale, electrode radii had a piecewise-linear relationship with the current thresholds for EFFECT (p < 0.005) and VF (p < 0.01). The liminal area determined by the piecewise-linear fit was 2.0 and 2.84 mm2 for EFFECT and VF, respectively. Above the liminal area, the threshold increased proportional to r1.25 and r0.95 (r = radius of electrode), for EFFECT and VF, respectively. Based on these experimental results, we present a theoretical framework to explain the electrode size-stimulation threshold variation for both low strength AC stimulation and VF initiation.
Keywords :
bioelectric phenomena; biological effects of fields; biomedical electrodes; electrocardiography; plethysmography; safety; stainless steel; AC cardiac stimulation; blood pressure collapse; current 25 muA to 3000 muA; electrode size-stimulation threshold; electrode surface area; frequency 160 Hz; frequency 20 Hz; frequency 40 Hz; frequency 60 Hz; frequency 80 Hz; isoflurane; lead II ECG; leakage currents; log-log scale; male breeder guinea pigs; medical devices; optical plethysmograph; piecewise-linear relationship; stainless steel rods; thoracotomy; tracheotomy; ventricular fibrillation; Biomedical electrodes; Biomedical monitoring; Biomedical optical imaging; Blood pressure; Electrocardiography; Fibrillation; Frequency; Leakage current; Piecewise linear techniques; Stimulated emission; Leakage currents; medical device safety; ventricular fibrillation; Animals; Differential Threshold; Electric Stimulation; Electricity; Equipment Design; Equipment Failure Analysis; Guinea Pigs; Male; Microelectrodes; Risk Assessment; Risk Factors; Ventricular Fibrillation;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2007.904528
Filename :
4303274
Link To Document :
بازگشت