DocumentCode
1136799
Title
Development of Shimming System for a Highly Sensitive NMR Spectrometer With a Superconducting Split Magnet
Author
Maki, K. ; Wakuda, T. ; Tsuchiya, M. ; Kido, S. ; Tsukamoto, H. ; Takeuchi, K. ; Okada, M. ; Kitaguchi, H.
Author_Institution
Hitachi Res. Lab., Hitachi Ltd., Hitachi
Volume
18
Issue
2
fYear
2008
fDate
6/1/2008 12:00:00 AM
Firstpage
844
Lastpage
847
Abstract
A highly sensitive NMR spectrometer with a superconducting split magnet and a solenoid-type probe coil is being developed. One of the most challenging tasks is to achieve high homogeneity of magnetic field in a measuring volume with this configuration, because inevitable structural asymmetry and limited production accuracy of the magnet cause larger inhomogeneity of magnetic field compared with a conventional solenoid-type NMR spectrometer. Numerical estimation revealed that field correction capability of conventional shim coils, i.e., superconducting solenoid and saddle coils placed outside main coils, is insufficient for the split NMR magnet. Therefore, we have designed and constructed superconducting shim coils with sufficiently high capability by adopting the following three novel ideas: (1) superconducting shim coils are placed not only outside main coils but also inside them in the vicinity of a measuring volume; (2) superconducting shim coils with ldquoperiodically wavyrdquo shape are utilized to correct non-axial inhomogeneous magnetic field; (3) the current of each superconducting shim coil is independently controlled to correct plural modes of magnetic field simultaneously.
Keywords
NMR spectrometers; superconducting coils; superconducting magnets; NMR spectrometer; current; saddle coils; shimming system; solenoid-type probe coil; superconducting split magnet; Linear programming; nuclear magnetic resonance; superconducting magnets;
fLanguage
English
Journal_Title
Applied Superconductivity, IEEE Transactions on
Publisher
ieee
ISSN
1051-8223
Type
jour
DOI
10.1109/TASC.2008.921881
Filename
4493380